The processing must be checked
regularly by the operator in order to
detect any foaming.
~
To avoid damage to the
washer-disinfector and the associated
accessories by the effect of process
chemicals, introduced contamination or
their interaction, the instructions in the
chapter "Chemical Process
Technology" must be followed.
~
Even when a chemical additive (e.g.
a cleaning agent) is recommended, the
washer-disinfector manufacturer takes
no responsibility for the effect of such
chemicals on the labware/instruments
involved.
Note that formulation changes, storage
conditions, etc., that are not disclosed
by the manufacturer may adversely
affect the cleaning results obtained.
~
When using additives and special
products, always follow the
manufacturer's instructions. To avoid
material damage and the possibility of
violent chemical reactions (e.g.
hydrogen gas reaction), always comply
with the manufacturer's intended
application(s) for a given cleaning
agent.
~
For critical applications, where very
high processing requirements have to
be met, it is strongly recommended that
all process-related factors (processing
chemicals, water quality, etc.) be
discussed in advance with Miele.
Safety Instructions and Warnings
~
In addition, for applications that
demand especially stringent cleaning
and rinsing results, (e.g., chemical
analysis) the operator must ensure that
quality control occurs on a regular
basis, to meet the standards involved.
~
Only use mobile units, baskets, and
inserts as intended by the
manufacturer.
Hollow items must be positioned for full
exposure to rinse water.
~
Empty all liquids from containers
and hollow items before loading them
into the washer-disinfector.
~
Ensure that no acid or solvent
residues, especially hydrochloric acid
or chloride solutions, get inside the
wash chamber. Never place corroding
ferrous materials in the
washer-disinfector. Only trace amounts
of any organic solvents should be
present in any soiling.
~
To avoid corrosive damage, make
sure the stainless steel housing does
not come into contact with
solutions/steam containing hydrochloric
acid.
~
After any work on the water supply
system, the water lines to the
washer-disinfector must be bled.
Otherwise some components of the
washer-disinfector may be damaged.
~
Please be sure to follow the
installation instructions in this manual
and in the installation manual.
9