Intended Use
For re-usable items to be safe for
continued use, the cleaning result must
meet certain minimum standards for
disinfection, sterilization and residue
levels.
Re-usable medical products are best
cleaned with the DES-VAR-TD process
or, as appropriate, with the
ORTHOVARIO or OXIVARIO processes.
Also important for acceptable cleaning
results is the use of suitable loading
accessories (mobile units, modules,
inserts). See the section on
"Applications" for examples.
The washer-disinfector is designed to
rinse with tap water or processed water
(e.g., distilled water, pure water,
deionized water, demineralized water)
depending on the quality required for
your application(s). The water quality
selected is especially important for
labware intended for analytical uses.
Under EN ISO 15883-1
(CAN/CSA-Z15883), this
washer-disinfector is suitable for
process validation.
12

User profiles

Daily-routine operators
For daily routine tasks, the operating
personnel must be instructed in the
basic functions and the loading of the
washer-disinfector and should receive
regular training. Basic knowledge of the
machine-based processing of medical
devices is required.
Daily routine work takes place on
operating levels A and C.
Daily-routine supervisors
Advanced knowledge of the
machine-based processing of medical
devices is required for more complex
tasks such as program interruption or
program cancellation.
These tasks take place on operating
level B.
Service tasks and administration
Additional expertise is required for
changing the processes or adjusting
the washer-disinfector, e. g. the
accessories that are used or the site
conditions.
Validations require special expertise in
the machine-based processing of
medical products, in the processing
technology and the applicable
standards and legislation.
Service tasks and validations are
carried out on operating level D.