Application; Protein Testing - Miele PG 8536 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

The washer-disinfector can be fitted
with various mobile units, adapted to
the type and shape of the items that
need to be cleaned and disinfected;
they can be equipped with a wide
variety of modules and inserts.
Mobile units, baskets and inserts
should be properly selected to
match the application involved.
On the following pages you will find
loading examples and additional
information.
Before EVERY program start,
conduct the following visual
inspection:
– Is everything correctly
sorted/loaded/connected for
cleaning?
– Are the spray arms clean and can
they rotate freely?
– Is the filter combination free of
debris (remove any coarse material
and clean the coarse, fine, flat and
micro-fine filters as needed)?
– Is the adapter properly connected to
the water supply for the spray arms
and nozzles?
– Are the process chemical dispensers
full (detergent, neutralizer, etc.)?
After EVERY program completion,
conduct the following visual
inspection:
– Inspect the labware or instruments
for cleanliness.
– Have any hollow items been
dislodged from their proper position
on the nozzles?
,
Any items that were dislodged
from the adapters during a program
cycle must be washed again.
– Are the cavities and channels in
hollow items cleaned through?
– Are the nozzles and connections
firmly attached to the mobile
units/inserts?

Protein testing

After washing, items used in medical
application should be checked at
random for proteins.

Application

33