Mobile units, baskets, and
inserts
To ensure that the mobile units,
baskets, and inserts function properly,
they must be checked daily. A checklist
is supplied with the washer-disinfector.
Check the following points:
– Are the rollers in proper condition,
and are they securely attached to the
mobile unit/insert?
– Is the mobile unit connection
correctly set and firmly screwed on?
– For mobile units in the modular
system, are the caps in the module
connection working properly?
– Are all spray jets, spray sleeves, and
hose adapters securely attached to
the mobile unit/insert?
– Are all spray jets, spray sleeves, and
hose adapters unclogged so that
wash water can flow through?
– Are all caps, covers, and fasteners
securely attached to the spray
sleeves?
Maintenance of the washer-disinfector
if applicable:
– Do the spray arms rotate freely?
– Are the spray jets clogged? See the
section on "Maintenance measures /
Cleaning the spray arms".
– Are the magnets in place on the
spray arm ends?
– Are the magnets on the spray arms
free of attached metal parts?
– Is the magnetic strip of the automatic
mobile unit recognition feature free of
attached metal objects?
– Are the screws in the magnet rails for
Automatic Mobile Unit Recognition
tightly secured?
Mobile units, modules and inserts
must also undergo routine
maintenance within the framework of
periodical maintenance of the
washer-disinfector every 1000
operating hours or at least once
per year.
See Maintenance of the
washer-disinfector.