Security 
Settings parameters are divided into four groups regarding security level: 
1 At the lowest level (PL0), where a password is not required, parameters of LCD can be set: language, 
contrast and LCD back light. 
2 At the first level (PL1), settings of a real time clock (MC350 and MC350H) can be changed and energy 
meters and MD can be reset. 
3 At the second level (PL2), the access to all data that are protected with the first level (PL1) and setting of 
all other parameters in the »SETTINGS« menu are available. 
4 A backup password (BP) is used if passwords at levels 1 (PL1) and 2 (PL2) have been forgotten, and it is 
different for each device (depending on a serial number of the meter). The BP password is available in 
the user support department in ISKRA, d.d., and is entered instead of the password PL1 or/and PL2. Do 
not forget to state the meter serial meter when contacting the personnel in ISKRA, d.d.. 
 
 PLEASE NOTE 
A  serial  number  of  device  is  stated  on  the  label,  LCD  (see  example  below)  and  is  also 
accessible with MiQen software. 
 
 
The access to the device serial number via a keyboard 
Example: 
 Main menu  Info OK  
 
Password setting     
A password consists of four letters taken from the British alphabet from A to Z. When setting a password, only 
the letter being set is visible while the others are covered with *. 
A password of the first (L1) and the second (L2) level is entered, and time of automatic activation is set. 
 Main menu  Settings  Security  Password level 1 / Password level 2 / Password lock time 
Password modification     
A password is optionally modified; however, only that password can be modified to which the access is unlocked 
at the moment. 
 Main menu  Settings  Security  Password level 1 / Password level 2 
Password disabling     
A password is disabled by setting the "AAAA" password.  
 Main menu  Settings  Security  Password level 1 / Password level 2   "AAAA" OK 
 
 PLEASE NOTE 
A factory set password is "AAAA" at both access levels (L1 and L2). This password does not 
limit access. 
 
 