Prog Menu; Figure 12 Main Menu - Prog Menu - Fluke 914 Series Manual

Hide thumbs Also See for 914 Series:
Table of Contents

Advertisement

914X Field Metrology Wells

Prog Menu

4.2
Prog Menu
F1
TEMP SETUP
F1
RUN PROGram
F2
RAMP/SOAK
F1
F3
SWITCH TEST
F1
F2
F4
TEST RESULT (-P only)
F1
F2
F3

Figure 12 Main Menu - Prog Menu

30
9142/9143/9144 MAIN MENU
F2
PROG MENU
SYSTEM MENU
TEST STATUS: <OFF, RUN> <select> Changes status of current test
RUN TEST: <RAMP/SOAK, SWITCH TEST> <select> Runs selected test
RECORD DATA: <YES, NO> <select> (-P only) Records data for ramp/soak or switch tests.
TEST ID: < Text, 16 char, English ASCII only > <edit> (-P only) Alpha-numeric identification for test
Number O.f SETPOINTS: <1-8> <select>
SOAK TIME: <minutes> <edit> Length of time in minutes at which the Field Metrology Well remains at the set point temperature
Number O.f CYCLES: <1-999> <select> Number of times the Field Metrology Well repeats the ramp/soak program
DIRECTION: < , > <select>
PASS TOLERANCE±: <edit> (-P only) Pass/Fail criteria for the ramp/soak program
SET POINTS
SETPOINT 1: <edit>
SETPOINT 2: <edit>
SETPOINT 3: <edit>
SETPOINT 4: <edit>
SETPOINT 5: <edit>
SETPOINT 6: <edit>
SETPOINT 7: <edit>
SETPOINT 8: <edit>
TEST METHOD: <AUTO, MANUAL> <select>
AUTO TEST
SWITCH TEMPerature: <nominal temperature> <edit> Run automated 3-cycle test based on the nominal temperature of the switch
MANUAL TEST (Run automated test based on test parameters set below)
UPPER TEMPerature: <edit> Upper temperature range in which encompasses a change of state for the temperature switch
LOWER TEMPerature: <edit> Lower temperature range in which encompasses a change of state for the temperature switch
APPROACH LIMIT: <edit> Temperature parameter in relation to the upper and lower temperature in which the system scan rate will be used
Number O.f CYCLES: <1-100> <edit> Number of times the Field Metrology Well repeats the manual test program
VIEW TESTS
DISPLAY TEST: <1-16> <select> Displays test results
DATE: <view only>
TIME: <view only>
PRINT TESTS
PRINT TESTS: < NO, YES> <select> Print all test data to RS-232 port
ERASE TESTS
ERASE TESTS: <NO, YES> <select> Erases all test data
F3
F4
INPUT SETUP (-P only)

Advertisement

Table of Contents
loading

This manual is also suitable for:

9143914291449143-dw-156

Table of Contents