Troubleshooting - Labnet BioPette PLUS BPP2 Instuction Manual

Up to 10000ml
Table of Contents

Advertisement

• Turn the key clockwise to reduce the aspirated volume,
or counter-clockwise to increase the volume. One full
turn of the calibration key changes the pipette aspira-
tion volume by the amount given in the table, (Fig. 4C),
• Take out the key and fix the pipetting pushbutton (Fig. 4D).
The pipetting pushbutton should be fixed by placing the
knob on the arbor first (Fig. 1A2) and then the pushbutton
(Fig. 1A1).
Determine the average aspirated volume. The average
volume should be within the permissible range given in
the table. If the volume exceeds the values stated, the
recalibration procedure should be repeated.
When pipetting liquids with physical properties conside-
rably different from those of water, follow the rules given
in section 5.

9 - TROUBLESHOOTING

If you notice an improper pipette operation, identify the
cause and eliminate the fault. To do this, follow the
instructions in the sequence provided. Replacement of
parts should be required only occasionally, and should
not occur under normal pipette use.
Droplets of liquid remain in the pipette tip.
• The tip is emptied too fast.
Decrease the speed of pressing the pipette push-
button.
• The tip wettability has increased due to extensive use.
Replace the tip with a new one.
Droplets of air appear in the liquid aspirated into the tip.
• The pipette tip immersion is too shallow.
Immerse the tip deeper according to the instructions.
• The pipette tip is incorrectly pressed onto the pipette
shaft.
Press the pipette firmly.
• The tip is damaged or worn out due to extensive use.
Replace the tip with a new one.
The pipette incorrectly aspirates the liquid or liquid
drops out from the tip.
• The pipette tip is incorrectly pressed onto the pipette
shaft.
Press the pipette tip firmly.
• The shaft nut is loose (Fig. 3F) in the models BPP2-
-BPP1000
11
Tighten the shaft nut.
• The sealing surface of the shaft is cracked or scored.
Remove the tip ejector. Unscrew the shaft nut,
inspect the shaft and the piston assembly. Replace
the damaged parts (see Section 12). When reasembling
the pipette, the nut should be hand tightened.
In the models BPP2, BPP10 and BPP20, the damage of
the shaft may also cause a damage of the piston
assembly. Replace the damaged parts (see Section 12).
When reasembling the pipette, the nut should be
hand tightened.
To remove the tip ejector in models BPP5000 and
BPP10000, remove the ejector pushbutton (Fig. 3N)
and using a screwdriver unscrew the tip ejector by
turning the screwdriver counter-clockwise.
• Damage to the piston or seal due to prolonged use
with the aggressive liquids.
Disassemble the pipette as described above.
Replace the piston, seal and O-ring (see Section 12).
Rinse the inside of the shaft in distilled water and dry.
Lubricate the seal and O-ring with the lubricant, that
has been included with each pipette.
The replacement of the piston requires conducting
of calibration procedure.
Note: The parts of BPP2 and BPP10 pipette should
be lubricated evenly with a minimum amount of
lubricant.
• The pipette is reassemled improperly.
Disassemble the pipette and reassemble it, observ-
ing the proper sequence of steps (Fig. 3).
• No lubricant on the sealing elements.
Remove the tip ejector. Unscrew the shaft nut,
remove the shaft, piston assembly, seal and O-ring.
Rinse the removed parts in distilled water and dry
thoroughly. Lightly lubricate the inside surfaces of
the seal and the O-ring with the included lubricant.
Reassemble the pipette in the reverse order.
• Contamination of the inside of the pipette caused by
extensive aspiration of chemically aggressive liquids or
because liquid got inside the pipette.
Remove the tip ejector. Unscrew the nut, remove
the shaft, piston assembly, seal and O-ring. Rinse
the removed parts with distilled water and dry thor
oughly. Lightly lubricate the inside surfaces of the
seal and the O-ring with the lubricant. Reassemble
the pipette in the reverse order.
12

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents