Data Transfer (Gasalert Extreme, Gasalertmicroclip, Gasalertmax Xt, And Gasalertquattro Only); Transferring Datalogs - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

MicroDock II
User Manual
Data Transfer
(GasAlert Extreme, GasAlertMicroClip, GasAlert-
Max XT, and GasAlertQuattro only)

Transferring Datalogs

The data transfer function transfers datalogs from the detector to the
base station's MMC/SD card. The data can then be imported into Fleet
Manager II.
Depending upon how many logs are stored on the MMC, the data trans-
fer may require several minutes to complete (1-2 minutes for
GasAlertMicroClip and GasAlertMax XT).
To transfer datalogs from the detector to the docking module and then to
the base station MMC/SD card, complete the following:
1. Activate the detector and wait until it is in normal operation.
Insert it into the docking module.
2. Activate the base station and wait until it is in normal operation.
C
3. Press
DATA TRANSFER on the docking module. The yellow
RUN LED flashes rapidly and the docking module transfers the
datalogs to the base station.
When the transfer process is complete, the docking module indi-
cates the status of the transfer as PASS or FAIL.
Successful Transfer: If the datalog transfer is successful, the
PASS LED lights green, the RUN LED flashes yellow slowly, and
the detector beeps.
Unsuccessful Transfer: If the datalog transfer is unsuccessful,
the FAIL LED lights red and the RUN LED flashes slowly.
72
If a failure occurs during the transfer of a datalog, typically some
of the data transfers to the MMC/SD card. To determine if any
data has been transferred, access the MMC data using Fleet
Manager II. Refer to the Fleet Manager II Operator's Manual.
If required, repeat step #3 to transfer the datalog.
4. After a successful transfer to the base station, the base station
transfers the datalog to the MMC/SD card. The RUN and the
PASS LEDs light on the docking module and the following screen
displays on the base station LCD.
When the datalog has successfully transferred from the base sta-
tion to the MMC/SD card, the following screen displays on the
base station LCD.

Hide quick links:

Advertisement

Table of Contents
loading