Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

2. Press
C
or
C
to move
3. Press
C
sel to access the inlet field.
on the left side of the inlet.
4. Press
C
or
C
to scroll to the desired inlet (1-5).
5. When the required inlet value displays, press
the selection and exit the field.
of the inlet, indicating the conflict has been resolved.
C
If
sel is not pressed within 30 seconds, the base station
automatically defaults to the inlet it has selected.
to the gas that is in conflict.
automatically displays
C
sel to confirm
no longer displays to the right
Note
Unsuccessful Conflict Resolution: If the conflict is not resolved by
selecting a different inlet, refer to the following:
Ensure that the selected inlet is correct.
Ensure the gas cylinder that is attached to the selected inlet is
correct.
Press
C
OK to accept the base station's default selection.

Not Found

If a value of 1 or more displays before Not Found, the base station is
unable to locate the required gas type for a specific sensor on the detec-
tor.
Press
C
cancel to exit the gas conflict screen and return to normal
operation. A bump test or calibration will not be performed.
To define the gas inlets, complete the following.
C
1. Press
more in the gas conflicts screen to access the gas
type/inlet screen to view which gas type(s) cannot be found.
MicroDock II
Gas Conflicts
57