MicroDock II
User Manual
Incorrect Inlet Setup
If an inlet is not setup correctly, a hyphen (
type.
, CO, LEL, and O
Example: SO
2
If a test gas fails, verify that the
•
gas cylinder is not empty,
•
connections are attached correctly,
•
inlets are setup correctly, and
•
gas concentration exceeds the low alarm setpoint (applicable to
GasAlertClip Extreme, GasAlert Extreme, and GasAlertMicro
only).
66
) displays beside the gas
gases were not detected.
2
Sensor Disabled
To prevent personal injury and/or property damage, enable or
replace the sensor immediately. Refer to the corresponding
detector user manual or technical reference guide for sensor
replacement procedures.
If a sensor is disabled, results from the sensor do not display on the base
station LCD.
displays where the gas type typically displays on the
LCD.
If a sensor is disabled, it is not recorded in the results log.
a Warning
Note