Inlet Select - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

MicroDock II
User Manual
C
5. Press
sel to confirm the selection and to access the format-
ting confirmation screen.
6. Press
C
Yes to initiate formatting. The following screen dis-
plays.
When formatting is complete, the LCD automatically returns to
the user options menu.
If No is selected, the LCD automatically returns to the user
options menu.
If there is a problem with the card, the following screen displays.
Press
C
OK to acknowledge and insert a new card and repeat
steps #4-6.
48
C
C
7. Press
or
Or
Press
C
exit to return to normal operation.
For additional information about the MMC/SD card, refer to
MMC/SD
Card.

Inlet Select

Applicable only to custom and multi-gas types.
Inlet Sel enables either the automatic (auto) or manual (manu) function
for selecting an inlet.
If the auto option is enabled, the base station automatically selects the
correct gas inlet for the bump test or calibration.
If manu option is enabled, the applicable test gas must be selected each
time a bump test or calibration is performed. If the manu option is
enabled, refer to
Order of Gases Applied for Bump Tests and Calibra-
tions
for important information.
To enable either the auto or manu option, complete the following:
1. From the user options menu, press
the Inlet Sel option.
to scroll to another user option.
Note
C
or
C
Base Station
to scroll to