Base Station Selects The Gas Inlet; User Selects Gas Inlet - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

MicroDock II
User Manual

Base Station Selects the Gas Inlet

To resolve a conflict automatically, complete the following steps.
1. Press
C
more in the gas conflicts screen to access the gas
type/inlet screen to view which gas is configured for more than
one inlet.
The gases display.
displays on the far right side on the row of the gas that is
setup for more than one inlet. The
the base station has selected as the optimum inlet.
2. Press
C
OK to accept the inlet that is selected and automati-
cally resume the bump test or calibration.
56
also indicates the inlet that
C
If
sel is not pressed within 30 seconds, the base station
automatically defaults to the inlet it has selected.

User Selects Gas Inlet

1. Press
C
more in the gas conflicts screen to access the gas
type/inlet screen to view which gas is configured for more than
one inlet.
The gases display.