The base station begins the calibration and displays the
docking module,
test being performed,
detector type, and
detector serial number.
The GasAlertMicro 5/PID/IR beeps and flashes two times when
calibration begins. The GasAlertMicroClip, GasAlertMax XT,
and GasAlertQuattro beep and flash once.
a Important
If the Automatic Datalog Download (GasAlertMicroClip,
GasAlertMax XT, GasAlertQuattro only) is enabled in Fleet
Manager II, data transfer automatically begins during the purge.
Do not remove the detector until the DATA TRANSFER PASS LED
lights, otherwise the transfer will fail.
5. If there is a gas conflict or a selected gas cannot be found, refer
to
Gas
Conflicts.
6. If there are no gas conflicts, the base station then applies the
gas.
Note
If more than one gas is being applied, the base station performs
a purge between each gas. Depending upon the gas type, the
purge time(s) will vary.
If Automatic Datalog Download is enabled in Fleet Manager
II, GasAlertMicroClip, GasAlertMax XT, and GasAlertQuattro
automatically deactivate after the transfer is complete. To
bypass the shutdown, remove the detector within 15 seconds
after the DATA TRANSFER PASS LED lights.
7. Refer to
Calibration
Results.
GasAlertMicro 5 IR with CO
If calibrating a GasAlertMicro 5 IR detector, the base station
automatically zeroes and then calibrates the CO
1. The base station zeroes the CO
only be zeroed using nitrogen (N
Ensure N
has been configured to an inlet. If required, refer to
2
Inlet
Setup.
2. The base station applies the gas.
MicroDock II
Calibration
Note
Sensor
2
Note
sensor first.
2
sensor. The CO
sensor can
2
2
).
2
69