Bump Test Results; Bump Pass; Bump Fail - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

Bump Test Results

After the bump test is complete, the base station displays the results of
the test.
Pass
Fail
Gas type not detected (H
H2S
example, this applies to all gases)
Sensor is disabled

Bump Pass

indicates that the detector passed the bump tests. The detector is
now ready for use. Press
C
S is used as an
2
OK to return to normal operation.

Bump Fail

If the detector failed a bump test,
Example: H2S
.
If the detector fails an audible or visual test,
cable test (AUD or VIS). For information regarding solutions, refer to
Troubleshooting.
If a gas type fails, calibrate the detector prior to use.
Confirm that the gas concentration exceeds the low alarm set-
point of the detector being tested (applicable to GasAlertClip
Extreme, GasAlert Extreme, and GasAlertMicro only).
Confirm the inlets are setup correctly.
MicroDock II
Bump Test
displays beside the applicable gas
displays beside the appli-