MODEL SOLU COMP II

5.7 SETTING A SECURITY CODE

5.7.1 Purpose.
This section describes how to set a security code. The security code prevents program and calibration settings
from accidentally being changed. Refer to Section 4.4 for additional information.
5.7.2 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
Outputs
Measurement
Temp
S S e e c c u u r r i i t t y y

5.8 NOISE REJECTION

5.8.1 Purpose.
For maximum noise rejection, the frequency of the ac power must be entered in the analyzer.
5.8.2. Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
O O u u t t p p u u t t s s
Measurement
T T e e m m p p
Security
N N o o i i s s e e R R e e j j e e c c t t i i o o n n
ResetAnalyzer
30
To store a number or setting, press ENTER.
Hold
1. Press MENU. The main menu screen appears. Choose Program.
Alarms
> > > >
2. Choose >>, then Security.
>>
3. Enter a three digit security code. The security code takes effect two min-
utes after the last key stroke.
4. The display returns to the security menu screen. Press EXIT to return to
the previous screen. To return to the main display, press MENU followed
by EXIT.
To store a number or setting, press ENTER.
Hold
1. Press MENU. The main menu screen appears. Choose Program.
Alarms
2. Choose >>.
>>
3. Choose >>.
>>
4. Choose Noise Rejection.
>>
5. Enter the mains frequency, 50 Hz or 60 Hz.
6. The display returns to the Noise Rejection screen. To return to the main
menu, press EXIT. To return to the main display, press MENU followed by
EXIT.
SECTION 5.0
PROGRAMMING THE ANALYZER