MODEL SOLU COMP II
5.4.3 Procedure: Configuring Alarms
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
Outputs
Measurement
Alarm Setpoints
A A l l a a r r m m C C o o n n f f i i g g u u r r e e
Alarm Config?
A A L L 1 1
AL2
AL1 S1 is for?
M M e e a a s s u u r r e e m m e e n n t t
AL3 is for?
Fault
S S e e n n s s o o r r 1 1
26
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
Hold
2. Choose Alarms.
A A l l a a r r m m s s
>>
3. Choose Alarm Configure.
4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).
AL3
5. For AL1 or AL2
a. Choose Measurement or Temp.
Temp
b. Choose High or Low.
c. Set the alarm Deadband.
6. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
7. For AL3
a. Choose Sensor1 (ozone) or Fault.
b. For sensor 1, choose Measurement or Temp.
c. Choose High or Low. Set the deadband.
d. Choosing Fault means AL3 will activate when a sensor or analyzer
fault exists. There is no user setting to make.
8. The display returns to the Alarm Configure? screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
SECTION 5.0
PROGRAMMING THE ANALYZER