SELECTING THE AMOUNT OF FILTERING

5.5.1 Purpose
This section describes how to change the amount of electronic filtering of the sensor current.
5.5.2 Definitions
INPUT FILTER. Before converting the sensor current to a ozone reading, the Solu Comp II applies an input filter.
The filter reduces noisy readings, but increases the response time. The level of filtering is selected by choosing
the amount of time required for the display to reach 63% of a step change.
5.5.4 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
Calibrate
P P r r o o g g r r a a m m
Display
Outputs
M M e e a a s s u u r r e e m m e e n n t t
Input filter?
63% in
0 0
05sec
28
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
Hold
2. Choose Measurement.
Alarms
>>
3. Choose the amount of filtering desired.