wtw Turb 555 Operating Manual page 68

Turbidity
Table of Contents

Advertisement

What to do if...
Measured values
that are obviously
too high
The sample does
not run out of the
pour-through vessel
Warning W-01
68
Cause
– Cuvette contaminated
– Cuvette scratched
– Cuvette misted up
– Air bubbles in the cuvette – Remove air bubbles
– Measuring instrument
has to be calibrated
Cause
– The end of the outlet
hose is placed too high
– There is a foreign body in
the pour-through vessel
Cause
– The light source is
defective
Turb 555 / Turb 555 IR
Remedy
– Clean the cuvette
– Replace the cuvette
– Regulate the temperature
of the sample before
measuring
– Run the instrument with
the dry air purge system
(see section 4.2.4)
– Calibrate the measuring
instrument
Remedy
– Place the end of the hose
lower than the measuring
instrument
– Remove the foreign body
Remedy
– Replace the lamp module
ba41113e07