wtw Turb 555 Operating Manual page 44

Turbidity
Table of Contents

Advertisement

Operation
Terminating the
calibration
prematurely
Calibration record
44
To terminate the calibration process before the scheduled
end, proceed as follows:
1 Wait until the current calibration step is finished and
the instrument asks you to insert the next calibration
standard.
8
2 Press the
lower display line.
1
3 Press the
The instrument returns to the measuring mode (
is displayed). If a printer or PC is connected and
switched on, a calibration record is automatically out-
put (see section 4.6.3 C
Note
If you have not carried out all calibration steps correctly, the
instrument cannot measure with the specified accuracy.
Note
You can switch off the instrument only after you have fin-
ished or terminated the calibration according to the program.
If a printer or PC is connected and switched on, a calibration
record is automatically output (sample printout, see section
4.6.3).
2
or
key until END is displayed in the
key.
ALIBRATION RECORD
Turb 555 / Turb 555 IR
a
).
ba41113e07
04/2009