wtw Turb 555 Operating Manual page 22

Turbidity
Table of Contents

Advertisement

Operation
Measuring units
Turb 555
Measuring units
Turb 555 IR
22
You can select the following measuring units (see section
4.5.2 S
):
ETTINGS
NTU = Nephelometric Turbidity Units (90°scattered light
measurement). Nephelometric turbidity unit according to
section 2130 of the "Standard Methods for the Examina-
tion of Water and Wastewater", 19th edition.
EBC = European Brewery Commission (90°scattered
light measurement). Nephelometric turbidity unit, used by
the European Brewery Commission
(Conversion: 0.245 EBC = 1 NTU).
NEPHELO (90°scattered light measurement)
Nephelometric turbidity unit, used in microbiological
research (Conversion: 6.7 Nephelo = 1 NTU).
NTU = Nephelometric Turbidity Units (90°scattered light
measurement). Nephelometric turbidity unit according to
section 2130 of the "Standard Methods for the Examina-
tion of Water and Wastewater", 19th edition.
FNU = Formazin Nephelometric Units (90°scattered light
measurement). Nephelometric turbidity unit. This unit is
only applicable if the instrument is calibrated with formaz-
in standards. It is used for measurements according to
DIN 27027 / ISO 7027 (Conversion: 1 FNU = 1 NTU).
EBC = European Brewery Commission (90°scattered
light measurement). Nephelometric turbidity unit, used by
the European Brewery Commission
(Conversion: 0.245 EBC = 1 NTU).
FAU = Formazin Attenuation Units (transmission mea-
surement). Formazin attenuation unit for measurements
according to DIN 27027 / ISO 7027 over 40 FNU.
Turb 555 / Turb 555 IR
ba41113e07
04/2009

Advertisement

Table of Contents
loading

This manual is also suitable for:

Turb 555 ir

Table of Contents