VERIFY OPERATION

Verification by LCD

Figure 5-9. Network Connection
LCD Display Messages
Field Communicator
Verification
5-14
PRELIMINARY
Operation can be verified in four ways, further described in this section:
1. At the device with the local display
2. Using the Field Communicator
3. In the Smart Wireless Gateway's integrated web interface
4. Using the AMS® Suite Wireless Configurator or AMS Device Manager
If the Rosemount 3308 was configured with the Network ID and Join Key, and
sufficient time has passed, the transmitter will be connected to the network. If
Network ID and Join Key were not configured, please see "Troubleshooting"
on page 5-16.
To verify operation at the device with the local display, confirm that the display
is showing a value. If the device is connected, the default LCD configuration
displays the primary variable (PV) value at the configured update rate.
illustrates the display messages that may be shown in the various stages of
the device's network connection. Refer to Section 6: Operation and
Maintenance for a full listing of error codes and other LCD messages. Press
the Diagnostic button to display the TAG, Device ID, Network ID, Network Join
Status, and Device Status screens.
N E T w K
S R C H N G
Searching for
Network
For HART Wireless transmitter communication, a Rosemount 3308 DD is
required. To obtain the latest DD, visit the Emerson Process Management
Easy Upgrade site at:
http://www2.emersonprocess.com/en-US/documentation/deviceinstallkits.
Connect the Field Communicator as shown on in Figure 5-2 on page 5-2 of
this document. Do not remove the Power Module. Removing the Power
Module may cause the device to drop off the network.
NOTE
In order to communicate with a Field Communicator, the device must be
powered by connecting the Power Module.
Enter the Field Communicator menu and use the Fast Key Sequence in
Table 5-3 to navigate to the Communications menu items. Select the
Communication Status parameter to verify operation. The remaining
Communications menu items provide additional network information.
n e t w k
n e t w k
L I M - O P
N E G O T
Joining
Connected with
Network
Limited Bandwidth
Reference Manual
00809-XXXX-4811, Rev AA
February 2012
n e t w k
O K
Connected