CONFIGURATION AND ADJUSTMENT

The EE650 as ordered is ready for use immediately and requires no configuration by the user. If required, the optional E+E Product
Configuration Adapter (EE-PCA) and the E+E Product Configuration Software (EE-PCS) can be used for changing the factory
setup as well as for adjusting of the air velocity measurement. Additionally, it allows for setting the display backlighting and
orientation.
PC
For product data sheets EE-PCS and EE-PCA please see www.epluse.com.
The E+E Product Configuration Software (EE-PCS) is free and can be downloaded from www.epluse.com/configurator.
MOUNTING
DRillinG in The Wall of The DucT foR insTallinG The MounTinG flanGe
Air flow
The arrow engraved on the sensing head of EE650 indicates the direction of the air stream during factory adjustment. When
installing the EE650 probe, make sure that the arrow matches exactly the flow direction.
Air flow
The mounting flange allows for precise setting of the EE650 immersion depth in a duct. The entire sensing head must be in the air
flow to be measured.
coRRecT
Air flow
EE-PCA
>16
(0.63")
60
(2.4")
coRRecT
incoRRecT
Immersion depth = 30 % - 50 %
of the duct diameter
HA011062
V03
incoRRecT