Technical Support And Repairs; Warranty Period; Specifications - Dentsply Sirona Cavitron Plus Directions For Use Manual

Ultrasonic scaler
Table of Contents

Advertisement

Available languages

Available languages

hold Tap-On
Foot Pedal for 2 seconds. If blinking
TM
stops, system is ready for use. Discard old insert or
return if within warranty. If blinking remains, refer to
Section 11.2 Technical Support and Repairs to
have unit serviced as soon as possible.
Slow Blinking (1 blink per second)
-- The system is not operating within factory

specifications.

1. Remove insert.
2. Turn Main Power Switch OFF, (O) position. Wait five
seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to section 11.2
Technical Support and Repairs to have unit
serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1.
Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2.
Turn Main Power Switch to the OFF (O) position.
Allow system to cool for 10 minutes and turn system
to the ON (I) position. Verify light is not illuminated.
3.
If light is still illuminated, refer to section 11.2
Technical Support and Repairs to have unit serviced as
soon as possible.

11 2 Technical Support and Repairs

For technical support and repair assistance call DENTSPLY
Professional Cavitron Care
989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M.
(Eastern Time). For areas outside the U.S., contact your local
DENTSPLY Professional representative.
SECTION 12:

warranty Period

The Cavitron Plus Ultrasonic Scaler is warranted for TWO
YEARS from date of purchase. The Steri-Mate
enclosed with your system is warranted for SIX MONTHS from
date of purchase. Refer to the Warranty Statement Sheet
furnished with your system for full Warranty Statement and
Terms.
Factory Certified Service at 1-800-
SM
Handpiece
®
SECTION 13:
Specifications
Electrical Voltage Continuous (100-240 VAC)
Current
1.0 Amperes, Maximum
Phase
Single
Frequency
50/60 Hertz
Water Pressure 20 to 40 psig
(138 to 275 kPa)
Water Temperature < 25
Water Flow Rate Minimum Setting (CCW) < 15 ml/min
Maximum Setting (CW) > 55 ml/min
Weight
4.4 lbs (2 Kg)
Dimensions
Height: 6 in. (15,24 cm)
Width: 9.5 in. (24,13 cm)
Depth: 8 in. (20,32 cm)
Handpiece Cable length: 6.5 ft. (2.0 M)
Auxillary Footswitch Cable length: 8 ft. (2.4 M)
Water Supply Line length: 8 ft. (2.4 M)
Footswitch
Protection Class IPX1. Not for operating
theatres
Remote Communication
Frequency: 2405 to 2480 MHz
Power: < 1mW
Channels: 16
Operating Environment
Temperature: 15 to 40 Deg Celsius
(59 to 104 Deg Fahrenheit)
Relative Humidity: 30% to 75% (non-condensing)
Transport and Storage Conditions
Temperature: -40 to 70 Deg. Celsius
(-40 to 158 Deg. Fahrenheit)
Relative Humidity:
10% to 100% (non-condensing)
Atmospheric Pressure: 500 to 1060 hPa
18
o
o
C (77
F)
.