MAG 8000 & MAG 8000 CT
8.

Technical data

8.1 MAG 8000 & MAG
8000 CT
8. Technical data
Description
Meter
Accuracy MAG 8000
Standard calibration
Extended calibration
Accuracy MAG 8000 CT
Custody transfer
Media conductivity
Temperature MAG 8000
Ambient
Media
Storage
Temperature MAG 8000 CT
Ambient
Media
Storage
Enclosure
Approvals MAG 8000
Approvals MAG 8000 CT
Conformity
Sensor
Size, flange and pressure
range MAG 8000
EN 1092-1 (DIN 2501)
ANSI 16.5 Class 150 lb
AS 4087
Size, flange and pressure
range MAG 8000 CT
EN 1092-1 (DIN 2501)
ANSI 16.5 Class 150 lb
AS 4087
Max. excitation frequency
(Transmitter decide the selected 6.25 Hz for sensor size DN 25 ... 150 (1" ... 6")
excitation frequency)
Liner
Electrode and grounding
electrodes
SFIDK.PS.026.D9.02
Specification
± 0.4% of rate ± 2 mm/s
± 0.2% of rate ± 2 mm/s
OIML R49 for DN 50 ... DN 300, Class 1 and 2 with turn down
up to Q3/Q1 = 400 at Q2/Q1 = 1.6
MI-001 verification for DN 50 ...DN 300, Class 2 with turn
down ratio Q3/Q1 = 80, Q3/Q1 = 63 or Q3/Q1 = 25 at Q2/Q1 = 6.3
Clean water > 20 μs/cm
−20 ... +60 °C (−4 ... +140 °F)
0 ... 70 °C (32 ... +158 °F)
−40 ... +70 °C (−22 ... +158 °F)
−20 ... +60 °C (−4 ... +140 °F)
0.1 ... 30 °C (32 ... +70 °F)
−40 ... +70 °C (−22 ... +158 °F)
IP68/NEMA 6P rating; Cable glands mounted requires Sylgard
potting kit to remain IP68/NEMA 6P, otherwise IP67/NEMA 4 rating
is obtained; Factory mounted cable provides IP68/NEMA 6P rating
Drinking water approval
NSF 61 (cold water) USA
WRAS (BS 6920 cold water) UK
ACS Listed France
DVGW W270 Germany
Belaqua (B)
OIML R49 pattern approval
Drinking water approval
NSF 61 (cold water) USA
WRAS (BS 6920 cold water) UK
ACS Listed France
DVGW W270 Germany
Belaqua (B)
OIML R49 pattern approval
MI-001 approval
(Number: DK-0200-MI-001-002
CEN EN 14154, ISO 4064
PED: 97/23EC
EMC: EN 61000-6-3, EN 61000-6-2, EN 61326-1
DN 25 and DN 40: PN 40
DN 50 ... 150: PN 16
DN 200 ... 600: PN 10 or PN 16
1" ... 2": 580 psi
2" ... 6": 230 psi
8" ... 24": 145 or 230 psi
DN 50 ... 600: PN 16
DN 50 ... 300: PN 16
2" ... 12": 230 psi
DN 50 ... 300: PN 16
3.125 Hz for sensor size DN 200 ... 600 (9" ... 24")
EPDM
Hastelloy C276
or
DK-0200-MI-001-004)
43