Verification; Mag 8000 Ct Sealing - Siemens SITRANS F M MAG 8000 Operating Manual

Battery operated water meter
Hide thumbs Also See for SITRANS F M MAG 8000:
Table of Contents

Advertisement

MAG 8000 & MAG 8000 CT
6.

Verification

6.1 MAG 8000 CT sealing

6. Verification
SFIDK.PS.026.D9.02
Verification mode increases the measurement
frequency to provide maximum measurements
per second. This function is especially useful to
minimize calibration rig time when validating
the flowmeter's accuracy. The frame around
the digits will blink slowly to indicate verification
mode is enabled. The maximum pulse rate on
output A is increased to 1 kHz and the pulse
width is set to 1 ms. After the verification mode,
the previous pulse setting is restored. Another
pulse width than 1 ms can be selected by
storing new pulse values. This setting remains
after the verification mode is finished.
Verification mode can be activated in two ways:
1) Via the Flow Tool (FT320)
2) By pushing the switch on the transmitter
PCB through the hole labeled "V" on the
front shield.
Verification mode stops automatically after 4
hours of operation. Verification mode can also
be aborted manually via the Flow Tool (FT320).
MAG 8000 CT user sealing marked with A.
MAG 8000 CT verification sealing marked with
A.
39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans f m mag 8000 ct

Table of Contents