Siemens SITRANS F M MAG 8000 Operating Manual page 44

Battery operated water meter
Hide thumbs Also See for SITRANS F M MAG 8000:
Table of Contents

Advertisement

MAG 8000 & MAG 8000 CT
(continued)
44
8. Technical data
Transmitter
Installation MAG 8000
Installation MAG 8000 CT
Enclosure
Cable entries
Display and key
Flow unit MAG 8000
Europe std.
US std.
Australian std.
Flow unit MAG 8000 CT
Europe std.
Digital output MAG 8000
Digital output MAG 8000 CT
SFIDK.PS.026.D9.02
Integral (compact) or remote with factory mounted cable in
5, 10, 20 or 30 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
Integral (compact) or remote with factory mounted cable in 5 or
10 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
MI-001 verified meters only Integral (compact) mounted.
Stainless steel top housing (AISI 316) and coated brass bottom.
Remote wall mount bracket in stainless steel (AISI 304).
2 x M20 (one gland for one cable of size 6 ... 8 mm
(0.02 ... 0.026 ft) is included in the standard delivery)
Display with 8 digits for main information.
Index, menu and status symbols for dedicated information
Key for toggling through the information and reset customer
totalizer and call-up function
Selectable default information and accessible menus:
- Operator
- Meter
- Service
- Data Logger
- Statistic and leakage (only Advanced version)
- Revenue and Tariffs (only Advanced version)
Totalized information can be displayed with 1, 2, 3 decimals or
automatic adjustment for maximum resolution
3
Volume in m
and flow rate in m
Volume in Gallon and flow rate in GPM
Volume in Ml and flow rate as Ml/d
Other units selectable:
3
Volume: m
x 100, l x 100, G x 100, G x 1000, MG,
CF x 100, CF x 1000, AF, Al, kl
3
3
Flow: m
/min, m
/d, l/s, l/min, GPS, GPH, GPD, MGD, CFS,
CFM, CFH
Other units are ordered from factory or manually configured onsite
by sticking a label on the display and changing the scaling factors
3
Volume in m
and flow rate in m
2 passive outputs (MOS), individual galvanically isolated
Maximum load ± 35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward – reverse – forward/net
– reverse/net
Output B function
Programmable as pulse volume (like output A), alarm or call-up
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
2 passive outputs (MOS), individual galvanically isolated
Maximum load ± 35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward
Output B function
– Alarm
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
3
/h
3
/h

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans f m mag 8000 ct

Table of Contents