Performing A Difference Measurement - wtw PhotoLab S6 Operating Instructions Manual

Hide thumbs Also See for PhotoLab S6:
Table of Contents

Advertisement

7. Method Parameters

7.1.2 Performing a difference measurement

Difference measuring is possible for some methods
(e.g. Iron II/III, Ca-/Mg Hardness).
For more information on this, see part, "Analysis
specifications".
Example:
Determination of iron (II) and iron (III).
– Change to measuring by pressing
18
– Enter method 106
– Confirm with
– Call up the citation menu item.
The current setting: Fe
– Using
– Confirm with
– Citation form Fe II, Fe III ∆ (S) is set.
4
.
– Start the 1st measurement by inserting cell 1.
1
.
3
scroll to Fe II, Fe III ∆
1
.