Measurements in normally charged surface water and waste water. Identification of the electrode A series number is impressed on the ORP electrode. Have this number ready in case of queries to the WTW service department. The series number provides the following information: Series number The number is impressed on the armoring.
Page 3
® SensoLyt MPP-A Pt Calibrating, measuring Calibrating, measuring Note See operating manual of the meter. Note Take off the watering cap when you want to calibrate or measure. Conversion to Meas standard hydrogen electrode with: = ORP, referring to the standard hydrogen electrode = Measured ORP Meas = Voltage of the reference system compared to the standard...
Page 4
® Maintenance (regeneration) SensoLyt MPP-A Pt Maintenance (regeneration) First-time For first-time activation, use the activating powder provided with the SORT/ activation during RH reagent set. Use a paper towel to gather a small amount of activating installation and as powder. Using moderate pressure, press the activating powder against the required moist (but not dripping) platinum electrode from two opposite sides and turn the electrode to and fro several times.
Page 5
(2) (see following figure). Maximum depth of immersion Note Detailed information on activating platinum electrodes, such as how to pro- duce the clorina solution, is given in the WTW application report entitled R . The application report is included in the GENERATING ELECTRODES SORT/RH reagent set.
Page 6
® Cleaning SensoLyt MPP-A Pt Cleaning Cleaning Contamination Cleaning procedure Gross contamination at the junction Carefully brush off contamination under running water using a soft toothbrush Fat, oil, protein-containing coatings Remove with household washing- and similar substances up liquid Accessories Description Model Order no.
® SensoLyt MPP-A Pt Technical data Technical data General features ORP electrode Platinum ring Reference electrolyte Polymer electrolyte Shunt conduction element Ag/AgCl Junction 2-hole junction Temperature sensor Integrated NTC 30 (30 kat 25 °C / 77 °F) Measurement and ORP application range -2000 ...