Tektronix AFG3000 Series Service Manual page 98

Arbitrary/function
Hide thumbs Also See for AFG3000 Series:
Table of Contents

Advertisement

Performance Verification
AFG3011
Function
Square
Square
AFG302xB
Function
Square
Square
AFG310x
Function
Square
Square
AFG325x
Function
Square
Square
4-36
4. Verify that the rise/fall time of the square waveform on the oscilloscope at each
amplitude is equal to or less than the limit specified in the following tables. Use
10
90% reference level for rise/fall time measurement.
Frequency
Offset
2.00 MHz
0.0 V
2.00 MHz
0.0 V
Frequency
Offset
10.00 MHz
0.0 V
10.00 MHz
0.0 V
Frequency
Offset
10.00 MHz
0.0 V
10.00 MHz
0.0 V
Frequency
Offset
10.00 MHz
0.0 V
10.00 MHz
0.0 V
5. (AFG3xx2 only) Repeat steps 2 through 4 for the channel 2 output.
AFG3000 Series Arbitrary/Function Generators Service Manual
Oscilloscope
Amplitude
Vertical
1.0 V
200 mV/div
pp
10.0 V
200 mV/div
pp
with x10 attenuator
Oscilloscope
Amplitude
Vertical
1.0 V
200 mV/div
pp
10.0 V
200 mV/div
pp
with x10 attenuator
Oscilloscope
Amplitude
Vertical
1.0 V
200 mV/div
pp
10.0 V
200 mV/div
pp
with x10 attenuator
Oscilloscope
Amplitude
Vertical
1.0 V
200 mV/div
pp
5.0 V
1.0 V/div
pp
Measurement
Horizontal
Limit
≤ 50 ns
20 ns/div or
25 ns/div
≤ 50 ns
20 ns/div or
25 ns/div
Measurement
Horizontal
Limit
≤ 18 ns
5 ns/div
≤ 18 ns
5 ns/div
Measurement
Horizontal
Limit
≤ 5 ns
2 ns/div
≤ 5 ns
2 ns/div
Measurement
Horizontal
Limit
≤ 2.5 ns
1 ns/div
≤ 2.5 ns
1 ns/div