Tektronix AFG3000 Series Service Manual page 81

Arbitrary/function
Hide thumbs Also See for AFG3000 Series:
Table of Contents

Advertisement

AFG3000 Series
Performance
Rise-Fall Time
AFG325x
CH1 Rise Time Amplitude: 1.0 V
Fall Time Amplitude: 1.0 V
Rise Time Amplitude: 5.0 V
Fall Time Amplitude: 5.0 V
CH2 Rise Time Amplitude: 1.0 V
Fall Time Amplitude: 1.0 V
Rise Time Amplitude: 5.0 V
Fall Time Amplitude: 5.0 V
AFG3000 Series Arbitrary/Function Generators Service Manual
Test Record
Minimum
- - - - -
pp
- - - - -
pp
- - - - -
pp
- - - - -
pp
- - - - -
pp
- - - - -
pp
- - - - -
pp
- - - - -
pp
Test result
Performance Verification
Maximum
2.5 ns
2.5 ns
2.5 ns
2.5 ns
2.5 ns
2.5 ns
2.5 ns
2.5 ns
4-19