Tektronix AFG3000 Series Service Manual page 84

Arbitrary/function
Hide thumbs Also See for AFG3000 Series:
Table of Contents

Advertisement

Performance Verification
4-22
4. Set up the arbitrary/function generator using the following steps:
Select menu
Setting
Function
Sine
Frequency
1.000000 kHz
Amplitude Units
V rms
The voltage unit should be
matched to that of DMM.
Amplitude
0.030 Vrms
Channel 1 Output On
5. Verify that each amplitude measurement is within the range specified in the
following tables.
AFG3011
Function
Frequency
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
AFG302xB and AFG310x
Function
Frequency
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
Sine
1.000 kHz
AFG3000 Series Arbitrary/Function Generators Service Manual
Operation
Sine (front)
Frequency/Period (front)
Top Menu > Amplitude/Level Menu > -more-
> Units > V rms
Amplitude/High (front)
On (front)
Measurement
Amplitude
30.0 mVrms
mVrms
300.0 mVrms
mVrms
800.0 mVrms
mVrms
1.500 Vrms
Vrms
2.000 Vrms
Vrms
2.500 Vrms
Vrms
3.500 Vrms
Vrms
Measurement
Amplitude
30.0 mVrms
mVrms
300.0 mVrms
mVrms
800.0 mVrms
mVrms
1.500 Vrms
Vrms
2.000 Vrms
Vrms
2.500 Vrms
Vrms
3.500 Vrms
Vrms
Range
(30.0 × CF ± 1.307) mVrms
(300.0 × CF ± 6.71) mVrms
(800.0 × CF ± 16.71) mVrms
(1.500 × CF ± 0.0307) Vrms
(2.000 × CF ± 0.0407) Vrms
(2.500 × CF ± 0.0507) Vrms
(3.500 × CF ± 0.0707) Vrms
Range
(30.0 × CF ± 0.654) mVrms
(300.0 × CF ± 3.35) mVrms
(800.0 × CF ± 8.35) mVrms
(1.500 × CF ± 0.0154) Vrms
(2.000 × CF ± 0.0204) Vrms
(2.500 × CF ± 0.0254) Vrms
(3.500 × CF ± 0.0354) Vrms

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Afg3021Afg3022Afg3101Afg3252Afg3102Afg3251