Tektronix AFG3000 Series Service Manual page 119

Arbitrary/function
Hide thumbs Also See for AFG3000 Series:
Table of Contents

Advertisement

AFG3000 Series Arbitrary/Function Generators Service Manual
Table 6-2: Internal inspection check list (cont.)
Item
Inspect for
Wiring and cables
Loose plugs or connectors. Burned, broken,
or frayed wiring.
Chassis
Dents, deformations, and damaged
hardware.
Interior Cleaning Procedure. Do the following steps to clean the arbitrary/function
generator interior:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and rinse with warm deionized water. (A cotton-tipped
applicator is useful for cleaning in narrow spaces and on circuit boards.)
If steps 1 and 2 do not remove all the dust or dirt, please contact Tektronix.
NOTE.
(See Contacting Tektronix at the front of this manual.)
Lubrication. There is no periodic lubrication required for the AFG3000 Series
arbitrary/function generator.
Maintenance
Repair action
Firmly seat connectors.
Repair or replace modules
with defective wires or
cables.
Straighten, repair, or
replace defective hardware.
6-5

Hide quick links: