Performance Verification
Calibration
4-2
Do the following steps to run the internal routines that confirm basic functionality
and proper adjustment:
1. Select Diagnostics in the Utility menu:
Utility (front-panel) > -more- (bezel) > Diagnostics/Calibration >
Execute Diagnostics
2. Wait until the test is completed.
3. Verify passing of the diagnostics.
If the diagnostics completes without finding any problems, the message
"PASSED" is displayed.
When an error is detected during diagnostic execution, the instrument displays
an error code. Error codes are described in Error codes on page 6-30.
4. Press any front-panel button to exit the diagnostics.
This procedure uses internal routines to verify that the instrument is operating
correctly.
Equipment required
None
Prerequisites
Do the following steps to run the internal routines that confirm basic functionality
and proper adjustment:
1.
Select Diagnostics in the Utility menu;
Utility (front-panel) > -more- (bezel) > Diagnostics/Calibration >
Execute Calibration
2. Wait until the test is completed.
3. Verify passing of the calibration.
If the calibration completes without any problems, the message "PASSED" is
displayed.
When an error is detected during calibration execution, the instrument displays
an error code. Error codes are described in Error Codes on page 6-30.
4. Press any front-panel button to exit the diagnostics.
AFG3000 Series Arbitrary/Function Generators Service Manual
Power on the instrument and allow a 20 minute warm-up before
doing this procedure.
An ambient temperature is between +20 °C and +30 °C.