AFG3000 Series Arbitrary/Function Generators Service Manual
AFG302xB and AFG310x
Function
Offset
DC
+ 5.000 Vdc
DC
0.000 Vdc
DC
– 5.000 Vdc
AFG3251 and AFG3252
Function
Offset
DC
+ 2.500 Vdc
DC
0.000 Vdc
DC
– 2.500 Vdc
6. (AFG3xx2 only) Repeat steps 3 through 5 for the channel 2 output.
Measurement
Range
(5.000 × CF ± 0.055) Vdc
Vdc
Vdc
(– 5.000 × CF ± 0.055) Vdc
Vdc
Measurement
Range
(2.500 × CF ± 0.030) Vdc
Vdc
Vdc
(– 2.500 × CF ± 0.030) Vdc
Vdc
Performance Verification
± 0.005 Vdc
± 0.005 Vdc
4-25