4

Quick Start

Caution: READ SECTION 6 ENTITLED BATH USE before placing the bath in
service. Incorrect handling can damage the bath and void the war ranty.

4.1 Unpacking

Unpack the Micro-Bath carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
4.2

Setup

Caution: DO NOT operate the instrument without fluid.
Place the calibrator on a flat surface with at least 6 inches of free space around the
instrument. Plug the power cord into a grounded mains outlet. Observe that the
nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe basket into the well. Fill the well with the appropriate fluid.
The set-point temperature and the number of and size of probes deter mine the fluid
level. Be sure to keep the fluid level an adequate distance below the top of the well
to prevent overflowing the fluid when the probes are in serted. For example, placing
200.05 oil at room temperature (25°C) into the bath and heating the unit to 126°C,
causes a 1-inch (2.54 cm) expansion of the fluid inside the well.
Keep the fluid level at least 1.9 cm (0.75 inches) below the top of the well at all
times. With the probe (probes) in the well fill the tank 3/4 full. Heat to the maximum
temperature of the fluid. Slowly fill the well to 2.54 cm (1 inch) be low the top of the
basket at the maximum temperature of the fluid, or to the second ring on the basket.
Shop for Fluke products online at:
7103 Micro-Bath
Transport/Pour Access Lid
Probe Basket
Stir Bar
Power Cord
Manual
Access Cover, if purchased
MyFlukeStore
www.
Quick Start
Unpacking
1.877.766.5412
.com
11