Controller Configuration
The proportional band width is set at the factory to about 5.0°C. The propor tional band
width may be altered by the user if he desires to optimize the con trol characteristics
for a particular application. The proportional band width is easily adjusted from the
front panel. The width may be set to discrete values in degrees C or F depending
on the selected units. The proportional band adjustment is be accessed within the
secondary menu. Press "SET" and "EXIT" to enter the secondary menu and show the
heater power. Then press "SET" to access the proportional band.
12.0 P
To change the proportional band press "UP" or "DOWN".
To store the new setting press "SET". Press "EXIT" to continue without storing the
new value.
9.8
Controller Configuration
The controller has a number of configuration and operating options and calibra tion
parameters which are programmable via the front panel. These are ac cessed from the
secondary menu after the proportional band function by pressing "SET". Pressing
"SET" again enters the first of three sets of configu ration parameters: operating
parameters, serial interface parameters, and cali bration parameters. The menus are
selected using the "UP" and "DOWN" keys and then pressing "SET". (See Figure 7
on page 32.)
9.8.1

Operating Parameters

The operating parameters menu is indicated by,
The operating parameters menu contains the High Limit and Stir Speed parameters.
38
Shop for Fluke products online at:
+
Access heater power in secondary menu
Heater power in percent
Access proportional band
Flashes "ProP" and the setting
ProP
Proportional band setting
5.0
New proportional band setting
4.0
Accept the new proportional band setting
Operating parameters menu
PAr
MyFlukeStore
www.
1.877.766.5412
.com