9.8.1.1
High Limit
The High Limit Parameter adjusts the upper set-point temperature. The factory default
and maximum temperature are set to 126°C. For safety, a user can ad just the HL down
so the maximum temperature set-point is restricted.
Press "SET" to enable adjustment of HL
Adjust the HL parameter using "UP" or "DOWN"
Press "SET" to accept the new temperature limit.
9.8.1.2
Stir Speed
The Stir Speed parameter adjusts stirrer motor speed. The factory default is 15.
To change the stir speed press "UP" or "DOWN".
Press "SET" to accept the new Stir Speed.
The stir motor speed needs to be varied for best stability. Table 4 on page 27 shows
nominal settings for several fluids.
9.8.2
Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
Shop for Fluke products online at:
High Limit parameter
HL
Flashes "HL" and then displays the setting
HL
Current HL setting
H=126
New HL setting
H=90
Flashes "Str Sp" and then displays the setting
Str SP
Current Stir Speed setting
0
New Stir Speed setting
16
Serial RS-232 interface parameters menu
SErIAL
MyFlukeStore
www.
Controller Operation
Controller Configuration
1.877.766.5412
.com
39