3

Specifications and Environmental Conditions

3.1 Specifications
Table 2 Specifications
Range
Accuracy
Stability
Uniformity
Resolution
Operating Temperature
Heating Time
Cooling Time
Well Size
Size
Weight
Display
Cooling
Fault Protection
Power
Safety
Fuse Rating

3.2 Environmental Conditions

Although the instrument has been designed for optimum durability and trou ble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and clean ing recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
Shop for Fluke products online at:
-30 to 125°C (-22 to 257°F)
±0.25°C
±0.03°C at -25°C (oil, 5010)
±0.05°C at 125°C (oil, 5010)
±0.02°C
0.01°C/F
5 to 45°C (41 to 113°F)
25°C to 100°C: 35 minutes
25°C to –25°C: 45 minutes
2.5" dia. x 5.5" deep (64 x 139 mm)
(access opening is 1.9" [48 mm] in diameter)
9" W x 13.2" H x 10.5" D (23 x 34 x 26 cm)
22 lb. (10 kg) with fluid
LED, Switchable °C or °F
Fan and Thermal Electric Devices (TED)
Sensor burnout and short protection
94 – 234 VAC (±10%), 50/60 Hz, 400 W
Overvoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01
250 V 5 A F(fast acting) NO USER SERVICEABLE PARTS
ambient temperature range: 5 – 45°C (41 – 113°F)
ambient relative humidity: maximum 80% for temperature <31°C, de creasing
linearly to 50% at 40°C
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters