7103 Micro-Bath User's Guide
Interface Commands

Table 5 Controller Communications Commands

Command Description
Command Format
Display Temperature
Read current set-point
s[etpoint]
Set current set-point to n
s[etpoint]=n
Read temperature
t[emperature]
Read temperature units
u[nits]
Set temperature units:
u[nits]=c/f
Set temperature units to Celsius u[nits]=c
Set temperature units to
u[nits]=f
Fahrenheit
Read scan mode
sc[an]
Set scan mode
sc[an]=on/off
Read scan rate
sr[ate]
Set scan rate
sr[ate]=n
Read hold
ho[ld]
Secondary Menu
Read proportional band setting
pr[opband]
Set proportional band to n
pr[opband]=n
Read heater power
po[wer]
(duty cycle)
Configuration Menu
Operating Parameters Menu
Read stirrer motor speed
mo[tor]
Set stirrer motor speed to n
mo[tor]=n
Read high limit
hl
Set high limit
hl=n
Serial Interface Menu
Read serial sample setting
sa[mple]
Set serial sampling setting to n
sa[mple]=n
seconds
Set serial duplex mode:
du[plex]=f[ull]/h[alf]
Set serial duplex mode to full
du[plex]=f[ull]
Set serial duplex mode to half
du[plex]=h[alf]
Set serial linefeed mode:
lf[eed]=on/of[f]
Set serial linefeed mode to on
lf[eed]=on
Set serial linefeed mode to off
lf[eed]=of[f]
Calibration Menu
Read R0 calibration parameter
r[0]
Set R0 calibration parameter to n r[0]=n
Read ALPHA calibration
al[pha]
parameter
Set ALPHA calibration parameter
al[pha]=n
to n
Read DELTA calibration
de[lta]
parameter
Set DELTA calibration parameter de[lta]=n
Read BETA calibration parameter be[ta]
Set BETA calibration parameter
be[ta]=n
Functions not on menu
Read firmware version number
*ver[sion]
Read structure of all commands h[elp]
Read all operating parameters
all
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and
linefeed. Then the value is returned as indicated in the RETURNED column.
46
Shop for Fluke products online at:
Command Example
Returned
s
set: 999.99 {C or F}
s=200.00
t
t: 999.99 {C or F}
u
u: x
u=c
u=f
sc
scan: {ON or OFF}
sc=on
sr
srat: 99.9 {C or F}/min
sr=1.1
ho
hold: open/closed,
99.9 {C or F}
pr
pb: 999.9
pr=8.83
po
po: 999.9
mo
mo: 99
mo=16
hl
hl:999
hl=90
sa
sa: 9
sa=0
du=f
du=h
lf=on
lf=of
r
r0: 999.999
r=100.324
al
al: 9.9999999
al=0.0038433
de
de:9.99999
de=1.3742
be
be:99.999
be=0.342
*ver
ver.9999,9.99
h
list of commands
all
list of parameters
MyFlukeStore
www.
Returned Example
Acceptable Values
set: 150.00 C
Instrument Range
t: 55.6 C
u: C
C or F
scan:ON
ON or OFF
srat:12.4C/min
.1 to 99.9
hold: open, 30.5 C
pb: 15.9
Depends on
Configuration
po: 1.0
mo: 15
0 to 40
hl:126
0–126
sa: 1
0 to 999
FULL or HALF
ON or OFF
r0: 100.578
90 to 110
al: 0.0038573
.002 to .005
de: 1.507
0–3.0
be:03427
–20 to 20
ver.7103,2.00
1.877.766.5412
.com