4 Mounting
Mount into the vessel
Mounting into the basin
18
4.6 Mounting steps with housing and thread
Fig. 9: Housing and thread
1
Housing
2
Seal
3
Thread
Mount VEGABAR 66 with housing and thread in the following way:
Weld the welded socket G1½ A or 1½ NPT to the vessel top
1
Insert the transmitter with connection tube or suspension cable
2
into the opening
Turn the thread with seal into the socket and tighten with SW 46
3
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Mount VEGABAR 66 with housing and thread in the following way:
Fasten the mounting bracket at the suitable height on the basin
1
wall
Information:
We recommend articles from the line of VEGA accessories:
Mounting bracket of stainless steel, article no. 2.21615
l
Counter nut of PP, article no. 2.10371
l
Insert the transmitter with connection tube or suspension cable
2
into the opening of the mounting bracket and counter nut
Fasten the counter nut to the thread with SW 46
3
Removal is carried out in reverse order.
Seal the 1½ NPT thread with teflon, hemp or a similar resistant material.
1)
VEGABAR 66 • 4 ... 20 mA/HART
1)