Dc Current Measurement; Resistance Measurement; Continuity Testing - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

DC Current Measurement

Range
±30.000 mA
±100.00 mA
Maximum Input: 110 mA
Maximum Burden Voltage: 420 mV at 22 mA
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
No Fuse
Normal mode rejection: 90 dB at 50 or 60 Hz nominal, and 60 dB at 1200 Hz and 2200 Hz nominal (HART signals)

Resistance Measurement

Range
Resolution
10.000 Ω
0.001 Ω
100. 00 Ω
0.01 Ω
1.0000 kΩ
0.1 Ω
10.000 kΩ
1 Ω
Open-circuit voltage: 5 V nominal
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)

Continuity Testing

Continuous tone
May or may not get tone
No tone
Find Quality Products Online at:
Resolution
1 Year
1 μA
0.01 % + 5 μA
10 μA
0.01 % + 20 μA
% of Reading + Floor
1 Year
0.05 % + 0.050 Ω
0.05 % + 0.05 Ω
0.05 % + 0.0005 kΩ
0.10 % + 0.010 kΩ
Tone
<25 Ω
25 to 400 Ω
>400 Ω
GlobalTestSupply
www.
Documenting Process Calibrator
Detailed Specifications
% of Reading + Floor
2 Year
0.015 % + 7 μA
0.015 % + 30 μA
Source
Current
2 Year
0.07 % + 0.070 Ω
3 mA
0.07 % + 0.07 Ω
1 mA
0.07 % + 0.0007 kΩ
500 μA
0.15 % + 0.015 kΩ
50 μA
Resistance
.com
sales@GlobalTestSupply.com
109

Advertisement

Table of Contents
loading

This manual is also suitable for:

754

Table of Contents