Supply Loop Power

The Product supplies loop power at 26 V dc through an
internal series resistance of 250 Ω. The setting supplies
sufficient current for two or three 4-20 mA devices on the
loop.
When you use loop power, the mA jacks are dedicated to
measuring the current loop. This means that the SOURCE
mA, measure RTD, and measure Ω functions are not
available (see Table 10.)
50
Find Quality Products Online at:
Connect the Product in series with the instrument current
loop as Figure 20 shows. To supply loop power:
1.
Push s for Setup mode.
Loop Power, Disabled is highlighted.
2.
Push  and  to select Disabled or Enabled.
3.
Push .
4.
Push the Done softkey. "LOOP" is shown on the
display when Loop Power is in operation.