Duration Limits - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

753/754
Users Manual
To log data:
1.
If necessary, push  for MEASURE mode.
2.
Push the More Choices softkey twice.
3.
Push the Log softkey.
4.
A list is shown; select a reading rate (1, 2, 5, 10, 20, 30,
or 60 readings per minute). Push  or  to select the
reading rate.
5.
Push .
6.
Push  to move the cursor to Duration.
7.
Use the numeric keypad to record the duration in
minutes, followed by . The maximum duration will
depend on the reading rate and how much memory is
available to log data.
84
Find Quality Products Online at:
Table 11 gives an estimate of the limits for duration,
assuming that no memory is being used for other purposes.
Readings/Minute
1
2
5
10
20
30
60
GlobalTestSupply
www.
Table 11. Duration Limits
Maximum
Approximate
Readings
Duration
8000
133 hours
8000
66 hours
8000
26 hours
8000
13 hours
8000
6 hours
7980
4 hours
7980
2 hours
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

754