DC Current Measurement
RANGE
RESOLUTION
1 µA
30 mA
10 µA
110 mA
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30V dc
Resistance Measurement
RANGE
RESOLUTION
11 Ω
0.001 Ω
110 Ω
0.01 Ω
0.1 Ω
1.1 kΩ
1 Ω
11 kΩ
Temperature Coefficient: (0.01% f.s. + 2 mΩ) /°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
MODEL 702
% OF READING +% OF FULL SCALE
1 YEAR
0.025% + 0.025%
0.025% + 0.03%
0.05% + 0.05%
0.07% + 0.07%
MODEL 702
% OF READING + Ohms
1 YEAR
0.05% + 0.05
0.075% + 0.05
0.05% + 0.05
0.075% + 0.05
0.05% + 0.5
0.075% + 0.5
0.1% + 10
0.1% + 10
Documenting Process Calibrator
% OF READING +% OF FULL SCALE
2 YEAR
1 YEAR
0.025% +0.025%
0.05% + 0.05%
% OF READING + Ohms
2 YEAR
1 YEAR
0.1% + 0.05
0.1% + 0.05
0.1% + 0.5
0.1% + 10
Specifications
MODEL 701
2 YEAR
0.025% + 0.03%
0.1% + 0.07%
MODEL 701
2 YEAR
0.15% + 0.05
0.15% + 0.05
0.15% + 0.5
0.15% + 10
89