GE akta pure Operating Instructions Manual page 82

Hide thumbs Also See for akta pure:
Table of Contents

Advertisement

4 Installation
4.2 Hardware installation
4.2.4 Install waste tubing
Step
3
4
82
Action
Cut the waste tubing to appropriate length. It is important that the tubing
is not bent and will not be submerged in liquid during the run.
Note:
If the tubing is too short, replace it with new tubing. Do not lengthen the tubing
as this might cause obstruction of the tubing.
Fasten all waste tubing securely to the waste vessel.