Power failure to...
Computer
Restart after emergency
shutdown or power failure
Follow the instructions to restart the system after an emergency shutdown or power
failure.
Step
1
ÄKTA pure Operating Instructions 29022997 AF
Action
Make sure that the condition that caused the emergency shutdown or
power failure is corrected.
will result in...
The UNICORN computer shuts down
•
The Power/Communication indicator (white)
•
on the Instrument control panel displays a
slowly flashing light.
The run is interrupted immediately
•
Data generated up to 10 seconds before the
•
power failure can be recovered
Note:
The UNICORN client may lose connection to the
instrument during a temporary overload of the
processor and display an error message. This
may appear as a computer failure. The run
continues and you can restart the UNICORN
client to regain control. No data will be lost.
2 Safety instructions
2.3 Emergency procedures
33