Power failure to...
Computer
Restart after emergency
shutdown or power failure
Follow the instructions to restart the system after an emergency shutdown or power
failure.
Step
1
ÄKTA pure Operating Instructions 29022997 AF
Action
Make sure that the condition that caused the emergency shutdown or
power failure is corrected.
will result in...
The UNICORN computer shuts down
The Power/Communication indicator (white)
on the Instrument control panel displays a
slowly flashing light.
The run is interrupted immediately
Data generated up to 10 seconds before the
power failure can be recovered
Note:
The UNICORN client may lose connection to the
instrument during a temporary overload of the
processor and display an error message. This
may appear as a computer failure. The run
continues and you can restart the UNICORN
client to regain control. No data will be lost.
2 Safety instructions
2.3 Emergency procedures
33