Monitor Status; Cell Diagnostics - ABB Aztec 600 Manual

Piston assembly single- and multi-stream colorimetric analyzers
Hide thumbs Also See for Aztec 600:

Advertisement

4.2 Initiating a manual piston clean / reset
IMPORTANT (NOTE)
Perform this procedure to clean the piston of any
solution / chemicals and reset it before returning it to
the factory for repair.
To initiate a manual clean / reset:
1. At the Indicator View screen, press the
Operate menu and use the
Diagnostics
2. Press the
key to select Diagnostics sub-menu options
and use the
and

Monitor Status

Cell Diagnostics

3. Press the
key to display Cell Diagnostics sub-menu
options:
IMPORTANT (NOTE)
If this procedure is selected while the analyzer is
operating, the following warning is displayed with
the Continue option active:
To exit press the
key.
key to display the
key to highlight Diagnostics:
keys to highlight Cell Diagnostics:
key, to proceed, press the
Aztec 600 | Piston assembly | Single- and multi-stream colorimetric analyzers | SI/ANAINST/001-EN
4. Use the
and
keys to navigate between fields and enter
data in the first 3 fields as shown below:
Fields
Selection / Data
Operation
Bring In
Volume
9 ml
Port
DI
Detector
Read-only
Voltage
(detector voltage [0 to 4095 mV] in real time).
Read-only
LED Current
(LED current in real time).
5. Use the
key to highlight the
the clean.
6. When the clean is complete, the
to the Operation field and select Reset Pump.
Fields
Selection / Data
Operation
Reset Pump
7. The piston can now be removed from the instrument and
returned to the factory for repair – refer to Section 4.3,
page 6.
button press
to initiate
and
keys to navigate
5