Process documentation

48
For data transfer between Profitronic system and an external report
printer or PC, a 5 m interface cable is included with the appliance.
The interface cable is wound up inside the machine and must only
connected by Miele Service.
The serial interface is RS 232 compatible. For the interface
configurations see "PC/Printer Function" in the Programming manual.
Various printers can be used as external printers:
– Miele PRT 100 report printer.
– Epson-compatible character set (font) (a list of suitable printers can
be obtained from Miele Service).
Pin configuration in the 9-pole sub-D connector on the interface
cable:
5 GND (ground)
3 TXD (transmit)
2 RXD (receive)
1-4-6 (linked)
7-8 (linked)
A standard null modem or laplink cable can be connected.
The maximum length of the extension cable for the printer/PC must
be 10 m.
If connecting a printer or PC, please note the following:
– Only use an industry-standard PC or printer (IEC 60950 certified).
– Select a printer or PC that is suited to the installation application.
For instructions on how to set the report printer functions, refer to the
section on PC/Print functions in the Programming Manual.
Various reports can be printed:
1. Program report
2. Optional function report
3. Wash protocol report
4. Fault protocol report