Warning and
Safety
Instructions
Connecting the
H
O
container
2
2
 These warning and safety instructions are in addition to those
given at the beginning of this booklet.
H
O
solution must only be used in special containers from "Dr.
2
2
Weigert", with the matching adapters.
Always comply with the safety instructions (material safety data
sheets) from the manufacturers of the process chemicals.
Be especially careful when handling H
chemical. Always comply with all applicable safety regulations!
Protective eyewear and gloves must be worn.
Empty containers must be disposed of in accordance with the
manufacturer's instructions.
H
O
solution must not be mixed with any other chemicals. This
2
2
could cause a violent chemical reaction, such as flash fire.
The connection hose for the H
black label.
The connection hose is supplied without an adapter as extraction
systems for hydrogen peroxide container will vary depending on
supplier.
Connect the appropriate adapter from the manufacturer to the
(black) connection line.
Connect the H
O
2
2
Start the DOS2-FILL service program.
 Unlike with other processing chemicals, the H
container must be fully emptied before it is exchanged for a new
one.
Do not replace the H
FILL DOS 2 CONTAINER message is displayed. After connecting
the new container you must run the DOS2-FILL service program.
If the CHECK DISPENSING SYSTEM 2 message appears you
should check the container as well as the dispensing system.
The program automatically pauses for this.
O
solution container is identified by a
2
2
container.
O
solution container with a new one until the
2
2
Optional feature
O
solution. It is a corrosive
2
2
O
solution
2
2