If a customer-specific calibration is required, this can be completed in the
menus described in the following. The calibration can be completed with the
connected probe or alternatively with a factory standard (e.g. Mahr 1283 WN
Factory Calibration, Order no. 5312833).
6.2
Customer calibration
6.2.1 Enabling customer calibration
–
Press the ON / OFF key
=>
Standard menu appears
–
Press u to select the System
menu
–
Press q to select the Customer
calibration submenu
ü
–
Press
to confirm
–
Press q to select the Enable
submenu
ü
–
Press
to confirm or Ó
cancel the selection
ü
=>
The
symbol indicates the
currently active calibration
–
Press the ON / OFF key to exit the
input menu.
36
to
Measurement
Display
Setup
Absolute mode
Customer calibration
Factory calibration
Factory settings
Keyboard-/ menu interlock
Info
ü
Ó
Measurement
Display
Setup
Customer calibration
Enable
Probe calibration
Factory standard calibration
Customer calibration Reset
ü
Ó
Measurement
Display
Setup
Absolute mode
Customer calibration
Factory calibration
Factory settings
Keyboard-/ menu interlock
Info
ü
Ó
Mahr GmbH • Millimar C1200
System
ü
q
p
System
q
p
System
ü
q
p